FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3972698 · Received August 1, 2014

Report

Report Number
2531779-2014-22049
Event Type
Malfunction
Date Received
August 1, 2014
Report Date
May 17, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 07/14/2014 WITH THE FOLLOWING FINDINGS: THERE WAS NO VISIBLE DAMAGE THE KEYPAD. THE KEYPAD BUTTONS WERE TESTED AND THE BUTTONS WERE FOUND TO BE RESPONDING PROPERLY TO PRESSES. THE KEYPAD WAS REMOVED AND NO CONTAMINATION WAS FOUND UNDER THE KEYPAD BUTTON CONTACTS. DURING TESTING, THE PUMP WAS FOUND TO BE LEAKING FROM A DAMAGED BOLUS BUTTON. THE PUMP WAS OPENED AND CORROSION WAS OBSERVED ON THE KEYPAD FLEX AND CONNECTION TO THE PRINTED CIRCUIT BOARD.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS ALLEGING THAT THE PUMP KEYPAD WAS NOT RESPONDING TO BUTTON PRESSES. THE REPORTER INDICATED THAT THE PUMP WAS STUCK ON THE HOME SCREEN BUT THE BUTTON PRESSES WERE CAUSING THE DISPLAY TO GET BRIGHTER BUT THERE WAS NOTHING ELSE. THE REPORTER INDICATED THAT THE PUMP HAD BEEN EXPOSED TO MOISTURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
450608 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 14 YR