FDA Adverse Event Malfunction Summary report: N

ANGLED ACET INSERTR

MDR report key: 3972618 · Received August 1, 2014

Report

Report Number
1818910-2014-24533
Event Type
Malfunction
Date Received
August 1, 2014
Date of Event
July 19, 2014
Report Date
July 19, 2014
Manufacturer
DEPUY INTERNATIONAL LTD. 8010379
Product Code
LXH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

THE INVESTIGATION CONFIRMED THAT THE ANGLED ACET INSERTER HANDLE HAD FRACTURED AS REPORTED. PREVIOUS INVESTIGATIONS HAVE FOUND THAT DESIGN CHANGE WAS IMPLEMENTED FOR THIS PRODUCT (ECO-321614) TO PREVENT THE ANTI-ROTATION PIN FROM LOADING THE PLASTIC HANDLE TO MINIMISE FURTHER FAILURES OF THE HANDLE CRACKING. DRA-001725 WAS CREATED TO INVESTIGATE THE USE OF RADEL IN INSTRUMENTS AND CONCLUDED THAT THE DATA SHOWED NO SYSTEMATIC FAILURES OF EXTRUDED RADEL FOR USE IN INSTRUMENTS. FOLLOWING THE DRA A HHE WAS COMPLETED (DVA-108291-HHE) AND CONCLUDED THAT THE RISK IS MEDIUM AND THAT NO FURTHER ACTION IS NECESSARY. THE COMPLAINT SHALL BE CLOSED WITH A JUSTIFIED CONCLUSION IT WILL BE ENTERED INTO THE COMPLAINT DATABASE AND MONITORED THROUGH TREND ANALYSIS. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.

Description of Event or Problem · 1

A LARGE PIECE OF THE HANDLE HAS BROKEN OFF AROUND THE HANDLE IMPACTING HEAD AREA. INSTRUMENT WAS FOUND IN A RETURNED LOAN KIT. THIS CASE HAS BEEN REPORTED BY THE LOAN KIT TECHNICIAN DURING LOAN KIT INSPECTION. THE INSTRUMENT HAD FURTHER BEEN ASSESSED BY DEPUY ENGINEERING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
451344 ANGLED ACET INSERTR HIP INSTRUMENT/TRIAL LXH DEPUY INTERNATIONAL LTD. 8010379 P28550001

Patients

Seq Age Sex Outcome Treatment
1