ANGLED ACET INSERTR
Report
- Report Number
- 1818910-2014-24533
- Event Type
- Malfunction
- Date Received
- August 1, 2014
- Date of Event
- July 19, 2014
- Report Date
- July 19, 2014
- Manufacturer
- DEPUY INTERNATIONAL LTD. 8010379
- Product Code
- LXH
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- PHYSICIAN
Narratives
THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
THE INVESTIGATION CONFIRMED THAT THE ANGLED ACET INSERTER HANDLE HAD FRACTURED AS REPORTED. PREVIOUS INVESTIGATIONS HAVE FOUND THAT DESIGN CHANGE WAS IMPLEMENTED FOR THIS PRODUCT (ECO-321614) TO PREVENT THE ANTI-ROTATION PIN FROM LOADING THE PLASTIC HANDLE TO MINIMISE FURTHER FAILURES OF THE HANDLE CRACKING. DRA-001725 WAS CREATED TO INVESTIGATE THE USE OF RADEL IN INSTRUMENTS AND CONCLUDED THAT THE DATA SHOWED NO SYSTEMATIC FAILURES OF EXTRUDED RADEL FOR USE IN INSTRUMENTS. FOLLOWING THE DRA A HHE WAS COMPLETED (DVA-108291-HHE) AND CONCLUDED THAT THE RISK IS MEDIUM AND THAT NO FURTHER ACTION IS NECESSARY. THE COMPLAINT SHALL BE CLOSED WITH A JUSTIFIED CONCLUSION IT WILL BE ENTERED INTO THE COMPLAINT DATABASE AND MONITORED THROUGH TREND ANALYSIS. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.
A LARGE PIECE OF THE HANDLE HAS BROKEN OFF AROUND THE HANDLE IMPACTING HEAD AREA. INSTRUMENT WAS FOUND IN A RETURNED LOAN KIT. THIS CASE HAS BEEN REPORTED BY THE LOAN KIT TECHNICIAN DURING LOAN KIT INSPECTION. THE INSTRUMENT HAD FURTHER BEEN ASSESSED BY DEPUY ENGINEERING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 451344 | ANGLED ACET INSERTR | HIP INSTRUMENT/TRIAL | LXH | DEPUY INTERNATIONAL LTD. 8010379 | P28550001 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |