FDA Adverse Event Malfunction Summary report: N

QUICKSET 1PC FLEX DRILL BIT 35

MDR report key: 3972612 · Received August 1, 2014

Report

Report Number
1818910-2014-24748
Event Type
Malfunction
Date Received
August 1, 2014
Date of Event
July 28, 2014
Report Date
July 28, 2014
Manufacturer
DEPUY ORTHOPAEDICS, INC. 1818910
Product Code
LXH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

(B)(4). EXAMINATION OF THE REPORTED DEVICE CONFIRMS THE REPORTED EVENT OF BIT BREAKAGE. A COMPLAINT DATABASE SEARCH HAS IDENTIFIED A TREND FOR THIS FAILURE WITHIN THE 2274 QUICKSET DRILL FAMILY. PREVIOUS INVESTIGATIONS HAVE DETERMINED THAT USE ERROR IS THE LIKELY ROOT CAUSE. AS A RESULT OF TRENDING HEALTH HAZARD EVALUATION, DVA-106292-HHE WAS CONDUCTED. THE INVESTIGATION DID NOT ESTABLISH THE NEED FOR CORRECTIVE ACTION. NO EVIDENCE WAS FOUND OF PRODUCT OR DESIGN ERROR AS A CONTRIBUTING FACTOR. SUBSEQUENT COMPLAINTS WILL BE MONITORED UNDER SEP 419 POST MARKET SURVEILLANCE. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY..

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Description of Event or Problem · 1

DRILL BIT BROKE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
451342 QUICKSET 1PC FLEX DRILL BIT 35 HIP INSTRUMENT/TRIAL LXH DEPUY ORTHOPAEDICS, INC. 1818910 PG235897

Patients

Seq Age Sex Outcome Treatment
1 53 YR