ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-22033
- Event Type
- Malfunction
- Date Received
- August 1, 2014
- Report Date
- July 26, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED, AND IT WAS EVALUATED BY PRODUCT ANALYSIS ON 09/10/2014 WITH THE FOLLOWING FINDINGS: VISUAL INSPECTION REVEALED THAT THE AUDIO BOLUS BUTTON COVER WAS MISSING, THUS THE BUTTON WAS DYSFUNCTIONAL; THE REPORTED EVENT WAS DUPLICATED. THE PUMP FAILED A LEAK TEST DUE TO AN AUDIO BOLUS BUTTON LEAK. THE PUMP CASE WAS REMOVED, AND EVIDENCE OF CORROSION WAS OBSERVED ON INTERNAL COMPONENTS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (ABB DMG PRIOR TCTL CNG W/MOIST) ISSUE. THE REPORTER ALLEGED THAT THE AUDIO BOLUS BUTTON WAS UNDER RESPONSIVE. REPORTEDLY, THE AUDIO BOLUS BUTTON COVER WAS MISSING/PEELING; MOISTURE/CORROSION WAS REPORTED TO BE EVIDENT IN THE AUDIO BOLUS BUTTON AREA. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 450478 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR |