ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-22010
- Event Type
- Malfunction
- Date Received
- August 1, 2014
- Report Date
- July 21, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 09/08/2014 WITH THE FOLLOWING FINDINGS: INVESTIGATORS WERE UNABLE TO DUPLICATE INTERMITTENT 'UP' BUTTON. TESTING CONFIRMED THE 'UP', 'DOWN', 'OK' AND 'CONTRAST' BUTTONS ARE RESPONSIVE TO BUTTON PRESSES AND ARE FUNCTIONAL. THE BUTTONS HAVE SPRING BACK AND CLICK. THE KEYPAD HAS NO VISIBLE SIGN OF DAMAGE. REMOVED KEYPAD COVER TO CHECK CONDITION OF THE BUTTON CONTACTS. NO CONTAMINATION OR ANY OTHER ANOMALY FOUND. PUMP WAS OPENED TO CHECK CONDITION OF THE KEYPAD FLEX AND CONNECTOR. THE KEYPAD FLEX IS FIRMLY SEATED IN CONNECTOR WITH NO SIGNS OF DAMAGE OR CONTAMINATION VISIBLE ON FLEX OR CONNECTOR. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 450210 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 22 YR |