FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3972486 · Received August 1, 2014

Report

Report Number
2032227-2014-05233
Event Type
Malfunction
Date Received
August 1, 2014
Date of Event
July 1, 2014
Report Date
July 1, 2014
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

INSULIN PUMP RECEIVED WITH UNRESPONSIVE BUTTONS DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARMS NOTED. CRACKED CASE AT DISPLAY WINDOW CORNER AND SEVERELY SCRATCHED LCD WINDOW NOTED.

Description of Event or Problem · 1

IT IS REPORTED THAT CUSTOMER RECIEVED BOTTON ERROR ALARM. CUSTOMERS BLOOD GLUCOSE LEVEL IS 246MG/DL. INSULIN PUMP IS BEING REPLACED. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
449960 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MMT-723RNAS

Patients

Seq Age Sex Outcome Treatment
1 78 YR