FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 3972452 · Received August 1, 2014

Report

Report Number
1416980-2014-24942
Event Type
Malfunction
Date Received
August 1, 2014
Date of Event
July 9, 2014
Report Date
July 9, 2014
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). (B)(6). EVALUATION SUMMARY: THE DEVICE WAS SERVICED ON-SITE BY A FIELD SERVICE TECHNICIAN. VISUAL INSPECTION, FUNCTIONAL TESTING, AND A REVIEW OF THE ALARM LOG WERE PERFORMED. FUNCTIONAL TESTING AND THE REVIEW OF THE ALARM LOG REVEALED THAT THE DEVICE HAD PRESENTED THE OPEN DOOR ALARM. VISUAL INSPECTION IDENTIFIED THAT THE DEVICE WAS MISSING A DOOR MAGNET. THE CAUSE OF THE REPORTED CONDITION WAS DETERMINED TO BE THE MISSING DOOR MAGNET. TO CORRECT THE CONDITION, THE DOOR MAGNET WAS REPLACED. THE DEVICE WAS SERVICED AND RESTORED TO A GOOD WORKING CONDITION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A FLOGARD INFUSION PUMP PRESENTED THE DOOR OPEN ALARM. THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS REQUESTED AND IS NOT AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
450072 FLOGARD PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1