PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 3004209178-2014-88742
- Event Type
- Malfunction
- Date Received
- August 1, 2014
- Date of Event
- July 1, 2014
- Report Date
- July 1, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP HAD CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS, MINOR SCRATCHES ON THE DISPLAY WINDOW, CRACKED RESERVOIR TUBE, MISSING END CAP STICKER AND CRACKED CASE NEAR DISPLAY WINDOW CORNERS NOTED DURING VISUAL INSPECTION.THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.
IT WAS REPORTED THAT THE INSULIN PUMP HAD A CRACK ON THE BATTERY COMPARTMENT. IT WAS NOTED THAT THE DRIVE SUPPORT CAP WAS DAMAGED. NO FURTHER INFORMATION REPORTED.