FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3972342 · Received July 31, 2014

Report

Report Number
2032227-2014-05189
Event Type
Malfunction
Date Received
July 31, 2014
Date of Event
July 1, 2014
Report Date
July 1, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA
Reporter Occupation
PATIENT FAMILY MEMBER OR FRIEND

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP RECEIVED WITH INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO MOISTURE DAMAGE FOUND ON ELECTRONIC ASSEMBLY/MOTOR. THE INSULIN PUMP HAD MINOR SCRATCHES ON LCD WINDOW, CRACKED DISPLAY WINDOW CORNER AND CRACKED RESERVOIR TUBE LIP.

Description of Event or Problem · 1

CUSTOMER'S MOTHER CALLED REPORTING HER DAUGHTER WAS AT CAMP AND SHE JUMPED IN THE POOL WHILE WEARING HER INSULIN PUMP. CUSTOMER'S MOTHER DID NOT KNOW CUSTOMER'S BLOOD GLUCOSE AT THE TIME. SHE WAS ADVISED THAT THE DEVICE NEEDS TO BE REPLACED, AND TO DISCONTINUE USE AND REVERT TO A BACK UP PLAN. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
447856 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-723LNAL

Patients

Seq Age Sex Outcome Treatment
1 12 YR