FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3972307 · Received July 31, 2014

Report

Report Number
3004209178-2014-88820
Event Type
Malfunction
Date Received
July 31, 2014
Date of Event
July 1, 2014
Report Date
July 2, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS UNABLE TO PRIME DURING THE PRIME/A33 TEST DUE TO PROTRUDED LOOSE DRIVE SUPPORT DISK. NO MOTOR ERROR ALARM AND MOTOR PASSED MOTOR TEST. THE INSULIN PUMP HAS MINOR SCRATCHED LCD WINDOW, CRACKED CASE NEAR DISPLAY WINDOW CORNERS, BATTERY TUBE THREADS CRACKED, BROKEN BELT CLIP SLOT, CRACKED RESERVOIR TUBE LIP AND MISSING END CAP STICKER.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER RECEIVED A MOTOR ERROR ALARM ON THE INSULIN PUMP. CUSTOMER STATED IT OCCURRED TWICE IN 12 HOURS. ADVISED THE CUSTOMER TO DISCONTINUE THE USE OF THE INSULIN PUMP AND REVERT TO BACK-UP PLAN. THE BLOOD GLUCOSE READING WAS 230MG/DL. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
447446 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722NAS

Patients

Seq Age Sex Outcome Treatment
1 46 YR