FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3972175 · Received July 31, 2014

Report

Report Number
2032227-2014-05126
Event Type
Malfunction
Date Received
July 31, 2014
Date of Event
July 1, 2014
Report Date
July 1, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

INSULIN PUMP RECEIVED WITH NO BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARM DURING TESTING. ALSO RECEIVED WITH CRACKED RESERVOIR TUBE LIP, MINOR SCRATCHED LCD WINDOW, BROKEN BELT CLIP SLOT AND SCRATCHED CASE.

Description of Event or Problem · 1

CUSTOMER REPORTED RECEIVING A LOW RESERVOIR ALARM ON THE INSULIN PUMP. HE STATED THAT THE ALARM COULD NOT BE CLEARED BY PRESSING THE ESCAPE OR THE ACT BUTTON. THE BLOOD GLUCOSE READING WAS 298 MG/DL. HE TREATED WITH MANUAL INJECTION. CUSTOMER REPORTED THAT THE BUTTONS ON THE INSULIN PUMP WERE NOT RESPONSIVE. NOTHING FURTHER WAS REPORTED

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
447481 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-722LNAS

Patients

Seq Age Sex Outcome Treatment
1 72 YR