FDA Adverse Event Injury Summary report: N

M2A-MAGNUM 42-50MM TPR INSRT-3

MDR report key: 3972172 · Received July 31, 2014

Report

Report Number
0001825034-2014-06684
Event Type
Injury
Date Received
July 31, 2014
Date of Event
April 17, 2014
Report Date
July 14, 2014
Manufacturer
BIOMET ORTHOPEDICS
Product Code
KWA
PMA / PMN Number
PK042037
Removal / Correction Number
N/A
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
ATTORNEY

Narratives

Additional Manufacturer Narrative · 1

CURRENT INFORMATION IS INSUFFICIENT TO PERMIT A CONCLUSION AS TO THE CAUSE OF THE EVENT. REVIEW OF DEVICE HISTORY RECORDS SHOW THAT LOT RELEASED WITH NO RECORDED ANOMALY RELATED TO THE EVENT. THERE ARE WARNINGS IN THE PACKAGE INSERT THAT STATE THAT THIS TYPE OF EVENT CAN OCCUR: UNDER POSSIBLE ADVERSE EFFECTS: "FRETTING AND CREVICE CORROSION MAY OCCUR AT INTERFACES BETWEEN COMPONENTS." AND "INTRAOPERATIVE OR POSTOPERATIVE BONE FRACTURE AND/OR POSTOPERATIVE PAIN." THIS REPORT IS BASED ON ALLEGATIONS SET FORTH IN PLAINTIFF¿S COMPLAINT AND THE ALLEGATIONS CONTAINED THEREIN ARE UNVERIFIED. THIS REPORT IS NUMBER 3 OF 3 MDRS FILED FOR THE SAME EVENT (REFERENCE 1825034-2014-05790 /-06683 /-06684).

Description of Event or Problem · 1

LEGAL COUNSEL FOR PATIENT REPORTED THAT PATIENT UNDERWENT LEFT TOTAL HIP ARTHROPLASTY ON (B)(6) 2009. SUBSEQUENTLY, LEGAL COUNSEL REPORTS THAT PATIENT¿S LEFT HIP WAS REVISED ON (B)(6) 2014 DUE TO PATIENT ALLEGATIONS OF PAIN. THIS REPORT IS BASED ON ALLEGATIONS SET FORTH IN PLAINTIFF¿S COMPLAINT AND THE ALLEGATIONS CONTAINED THEREIN ARE UNVERIFIED. OPERATIVE REPORT RECEIVED NOTED PATIENT UNDERWENT A LEFT HIP REVISION ON (B)(6) 2014 DUE TO PAIN. OPERATIVE REPORT NOTED THE PRESENCE OF SCAR TISSUE, A DEFICIENT MEDIAL WALL, THE HEAD/TAPER ADAPTER WERE COLD-WELDED ONTO THE STEM AND THE FEMUR FRACTURED DURING REMOVAL OF THE STEM. ALL COMPONENTS WERE REMOVED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
447480 M2A-MAGNUM 42-50MM TPR INSRT-3 PROSTHESIS, HIP KWA BIOMET ORTHOPEDICS N/A 810930

Patients

Seq Age Sex Outcome Treatment
1 59 YR Hospitalization| R