FDA Adverse Event Malfunction Summary report: N

FREESTYLE LITE

MDR report key: 3972141 · Received July 31, 2014

Report

Report Number
2954323-2014-00829
Event Type
Malfunction
Date Received
July 31, 2014
Date of Event
May 23, 2014
Report Date
May 23, 2014
Product Code
NBW
PMA / PMN Number
K092602
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION OF THE RETURNED PRODUCT DETERMINED THE CAUSE TO BE ISOLATED TO SOFTWARE CORRUPTION. ALTHOUGH GLUCOSE RESULTS MAY BE DELAYED, BLOOD GLUCOSE COULD BE DETERMINED BY ALTERNATE MEANS, INCLUDING USE OF ANOTHER BLOOD GLUCOSE METER, SEEING A PHYSICIAN (AS RECOMMENDED IN PRODUCT LABELING), OR BY SEEKING TREATMENT AT A HEALTH CARE FACILITY. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED.

Description of Event or Problem · 1

CUSTOMER INITIALLY REPORTED THAT THEIR ADC METER DID NOT TURN ON. THE PRODUCT WAS RETURNED AND INVESTIGATED. THIS MDR IS BEING SUBMITTED DUE TO RETURNED PRODUCT INVESTIGATION RESULTS. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
447940 FREESTYLE LITE BLOOD GLUCOSE MONITORING SYSTEM NBW

Patients

Seq Age Sex Outcome Treatment
1