ARTICULEZE M HEAD 36MM +12
Report
- Report Number
- 1818910-2014-24527
- Event Type
- Injury
- Date Received
- July 31, 2014
- Date of Event
- July 21, 2014
- Report Date
- August 25, 2014
- Manufacturer
- DEPUY INTERNATIONAL LTD. 8010379
- Product Code
- JDI
- PMA / PMN Number
- PK980513
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- PHYSICIAN
Narratives
THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
PATIENT WAS REVISED TO ADDRESS PAIN. DOI: 14 YRS AGO - DOR: (B)(6) 2014 (LEFT HIP). THE DEVICES ASSOCIATED WITH THIS REPORT WERE NOT RETURNED. PER WI-3430, A REVIEW OF THE DEVICE HISTORY RECORDS FOR THE PROVIDED PRODUCT AND LOT COMBINATIONS IS NO LONGER REQUIRED. REQUESTS FOR ADDITIONAL INVESTIGATIONAL INPUTS WERE MADE IN ACCORDANCE WITH WI-7915 APPENDIX A. NO ADDITIONAL INFORMATION WAS OBTAINED. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY PRODUCT CONTRIBUTION TO THE REPORTED EVENT WITH THE INFORMATION PROVIDED. BASED ON THE INABILITY TO IDENTIFY ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.
PATIENT WAS REVISED TO ADDRESS PAIN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 447486 | ARTICULEZE M HEAD 36MM +12 | HIP FEMORAL HEAD | JDI | DEPUY INTERNATIONAL LTD. 8010379 | YLN49 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 62 YR | Required Intervention |