COULTER LH 500 HEMATOLOGY ANALYZER
MDR report key: 3972115
·
Received July 31, 2014
Report
- Report Number
- 1061932-2014-01798
- Event Type
- Malfunction
- Date Received
- July 31, 2014
- Date of Event
- July 7, 2014
- Report Date
- July 7, 2014
- Manufacturer
- BECKMAN COULTER
- Product Code
- GKZ
- PMA / PMN Number
- K042724
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
THE BIOMEDICAL ENGINEER AT THE CUSTOMER'S FACILITY SERVICED THE INSTRUMENT. THE BIOMEDICAL ENGINEER CONFIRMED THE EVENT AND RESOLVED IT BY REPLACING WORN TUBING THROUGH PINCH VALVE PV37. (B)(4).
Description of Event or Problem · 1
CUSTOMER REPORTED A LEAK WHEN USING THE COULTER LH 500 HEMATOLOGY ANALYZER. THE LEAK WAS AN UNCONTAINED, 2-3 ML OF BLUE FLUID FROM THE INSTRUMENT, AND WAS IDENTIFIED AFTER PERFORMING AN INSTRUMENT SHUTDOWN PROCESS. THE OPERATOR WAS WEARING PERSONAL PROTECTIVE EQUIPMENT OF LAB COAT AND GLOVES WHEN THE LEAK WAS IDENTIFIED. THERE WAS NO BIOHAZARD EXPOSURE TO MUCOUS MEMBRANES OR CUTS. THERE WERE NO ERRONEOUS TEST RESULTS ASSOCIATED WITH THIS EVENT. THERE WAS NO DEATH, INJURY OR AFFECT TO USER OR PATIENT TREATMENT.