FDA Adverse Event Malfunction Summary report: N

COULTER LH 500 HEMATOLOGY ANALYZER

MDR report key: 3972115 · Received July 31, 2014

Report

Report Number
1061932-2014-01798
Event Type
Malfunction
Date Received
July 31, 2014
Date of Event
July 7, 2014
Report Date
July 7, 2014
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K042724
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE BIOMEDICAL ENGINEER AT THE CUSTOMER'S FACILITY SERVICED THE INSTRUMENT. THE BIOMEDICAL ENGINEER CONFIRMED THE EVENT AND RESOLVED IT BY REPLACING WORN TUBING THROUGH PINCH VALVE PV37. (B)(4).

Description of Event or Problem · 1

CUSTOMER REPORTED A LEAK WHEN USING THE COULTER LH 500 HEMATOLOGY ANALYZER. THE LEAK WAS AN UNCONTAINED, 2-3 ML OF BLUE FLUID FROM THE INSTRUMENT, AND WAS IDENTIFIED AFTER PERFORMING AN INSTRUMENT SHUTDOWN PROCESS. THE OPERATOR WAS WEARING PERSONAL PROTECTIVE EQUIPMENT OF LAB COAT AND GLOVES WHEN THE LEAK WAS IDENTIFIED. THERE WAS NO BIOHAZARD EXPOSURE TO MUCOUS MEMBRANES OR CUTS. THERE WERE NO ERRONEOUS TEST RESULTS ASSOCIATED WITH THIS EVENT. THERE WAS NO DEATH, INJURY OR AFFECT TO USER OR PATIENT TREATMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
447736 COULTER LH 500 HEMATOLOGY ANALYZER COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER NA NA