COULTER LH 500 HEMATOLOGY ANALYZER
Report
- Report Number
- 1061932-2014-01780
- Event Type
- Malfunction
- Date Received
- July 31, 2014
- Date of Event
- July 14, 2014
- Report Date
- July 14, 2014
- Manufacturer
- BECKMAN COULTER
- Product Code
- GKZ
- PMA / PMN Number
- K042724
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
A BECKMAN COULTER (BEC) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO ASSESS THE INSTRUMENT. THE FSE FOUND CRIMPED WASTE LINE TUBING AT THE NEEDLE CAUSED BY A LOOSE SCREW AT THE BOTTOM OF THE NEEDLE HOLDER. THE FSE TIGHTENED THE SCREW AND THE INSTRUMENT PERFORMED WITHOUT LEAKS. THE FSE VERIFIED THE INSTRUMENT AND THE INSTRUMENT WAS RETURNED TO NORMAL OPERATION. (B)(4).
THE CUSTOMER REPORTED APPROXIMATELY 5 MLS OF BLOOD LEAKED INSIDE THE COULTER LH 500 HEMATOLOGY ANALYZER ONTO THE COUNTER WHILE PERFORMING SAMPLES. THE LEAK WAS NOT CONTAINED WITHIN THE INSTRUMENT. THERE WAS NO EXPOSURE TO OPEN WOUNDS OR MUCOUS MEMBRANES. THE CUSTOMER WAS WEARING PERSONAL PROTECTIVE EQUIPMENT (PPE) CONSISTING OF GLOVES, GOGGLES AND A LAB COAT AT THE TIME OF THE OCCURRENCE. NO ERRONEOUS RESULTS WERE GENERATED IN CONNECTION WITH THE REPORTED EVENT. THERE WAS NO DEATH, INJURY, OR AFFECT TO USER OR TO PATIENT TREATMENT ATTRIBUTED TO THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 447455 | COULTER LH 500 HEMATOLOGY ANALYZER | COUNTER, DIFFERENTIAL CELL | GKZ | BECKMAN COULTER | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |