PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3972077 · Received July 31, 2014

Report

Report Number
2032227-2014-05077
Event Type
Malfunction
Date Received
July 31, 2014
Date of Event
May 21, 2014
Report Date
July 2, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. UNIT RECEIVED WITH CRACKED CASE ON DISPLAY WINDOW CORNERS, SCRATCHED LCD WINDOW, CRACKED RESERVOIR TUBE LIP, MISSING END CAP STICKER, AND MISSING ADDRESS/SERIAL NUMBER LABEL.

Description of Event or Problem · 1

IT WAS REPORTED THE CUSTOMER'S INSULIN PUMP ALARMED BUTTON ERROR AND IS NOT WORKING. CUSTOMER'S BLOOD GLUCOSE LEVEL WAS 132 MG/DL. CUSTOMER'S MOTHER STATED, THE CUSTOMER WAS AT SCHOOL AND THE NURSE CALLED HER TO INFORM HER THE CUSTOMER'S DEVICE WAS ALARMING. CUSTOMER'S MOTHER WAS ADVISED TO DISCONTINUE USE OF THE DEVICE AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
447156 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-523LNAH

Patients

Seq Age Sex Outcome Treatment
1 11 YR