FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3972073 · Received July 31, 2014

Report

Report Number
2032227-2014-05086
Event Type
Malfunction
Date Received
July 31, 2014
Date of Event
May 27, 2014
Report Date
July 2, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP PASSED THE REWIND, BASIC OCCLUSION TEST, OCCLUSION TEST, PRIME/A33 TEST, EXCESSIVE NO DELIVERY TEST, AND DISPLACEMENT TEST. NO MOTOR ERROR ALARMS NOTED. THE DEVICE WAS RECEIVED WITH INTERMITTENT BUTTONS DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARMS NOTED. THE DEVICE WAS RECEIVED WITH CRACKED CASE AT DISPLAY WINDOW CORNERS, RESERVOIR TUBE AND BATTERY TUBE THREADS. THE DEVICE WAS RECEIVED WITH SCRATCHED DISPLAY WINDOW AND WAS MISSING END CAP STICKER. THE MOTOR WAS TESTED OUTSIDE THE DEVICE AND PASSED.

Description of Event or Problem · 1

CUSTOMER REPORTED THE INSULIN PUMP ALARMED MOTOR ERROR AND NOTICED THE SCREEN WAS CRACKED. CUSTOMER'S BLOOD GLUCOSE LEVEL WAS 132 MG/DL. CUSTOMER DOES NOT RECALL ANY PREVIOUS SIGNIFICANT EVENT THAT MAY HAVE CAUSED THE DEVICE TO ALARM. CUSTOMER STATED THEY ARE UNABLE TO REWIND THE DEVICE. CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE DEVICE AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
448344 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-523LNAP

Patients

Seq Age Sex Outcome Treatment
1 46 YR