FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3972061 · Received July 31, 2014

Report

Report Number
2032227-2014-05088
Event Type
Malfunction
Date Received
July 31, 2014
Date of Event
June 2, 2014
Report Date
July 2, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH INTERMITTENT BUTTON RESPONSE AND BUTTON ERROR ALARM DUE TO CORRODED KEYPAD TRACES. THE DEVICE HAD CRACKED DISPLAY WINDOW CORNER, SCRATCHED LCD WINDOW, CRACKED BATTERY TUBE THREADS, CRACKED BELT CLIP SLOT AT BATTERY TUBE THREADS AREA, AND CRACKED RESERVOIR TUBE LIP.

Description of Event or Problem · 1

IT WAS REPORTED THE CUSTOMER'S INSULIN PUMP ALARMED BUTTON ERROR. CUSTOMER'S BLOOD GLUCOSE LEVEL WAS 405 MG/DL. CUSTOMER'S MOTHER REPORTED THAT HT CUSTOMER WAS SWEATING AS A SIGNIFICANT EVENT THAT MAY HAVE CAUSED THE DEVICE TO ALARM. CUSTOMER'S MOTHER WAS ADVISED THE CUSTOMER HAS TO DISCONTINUE USE OF THE DEVICE AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
448487 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-723LNAB

Patients

Seq Age Sex Outcome Treatment
1 16 YR