FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3972042 · Received July 31, 2014

Report

Report Number
2032227-2014-05062
Event Type
Malfunction
Date Received
July 31, 2014
Date of Event
July 1, 2014
Report Date
July 1, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE, THEREFORE, CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE BATTERY ON THE INSULIN PUMP IS NOT LASTING MORE THAN 12 HOURS. CUSTOMER STATED THAT SHE CHANGED THE BATTERY MULTIPLE TIMES AND THE SCREEN DISPLAY IS BLANK. DURING TROUBLESHOOT; IT WAS STATED THAT THERE IS CORROSION INSIDE THE BATTERY CAP AND THE CONTACTS ON THE BATTERY CAP ARE DAMAGED. IT WAS ALSO STATED THAT THE INSULIN PUMP IS ALARMING FOR HIGH BLOOD GLUCOSE AND SHE IS NORMALLY 5 MMOL/L. CURRENT BLOOD GLUCOSE IS 11 MMOL/L. THE BATTERY CAP IS BEING REPLACED. NOTHING FURTHER REPORTED.