FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 3972037
·
Received July 31, 2014
Report
- Report Number
- 2032227-2014-05047
- Event Type
- Malfunction
- Date Received
- July 31, 2014
- Date of Event
- April 14, 2014
- Report Date
- July 1, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BE
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE INSULIN PUMP WAS RECEIVED WITH A BUTTON ERROR ALARM AND NO BUTTON RESPONSE DUE TO MOISTURE-DAMAGED KEYPAD TRACES. THE UNIT HAD A SCRATCHED DISPLAY WINDOW, A SCRATCHED CASE, CRACKED BATTERY TUBE THREADS, A BROKEN BELT CLIP SLOT, AND A CRACKED RESERVOIR TUBE LIP.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DOWN ARROW BUTTON ON THE CUSTOMER'S INSULIN PUMP WAS NOT WORKING PROPERLY. THE CUSTOMER STATED THAT THE BUTTON WOULD NO LONGER DEPRESS. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 447115 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-754LWWS |