PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 3004209178-2014-88614
- Event Type
- Malfunction
- Date Received
- July 31, 2014
- Date of Event
- July 2, 2014
- Report Date
- July 2, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- OTHER
Narratives
NO BUTTON ERROR ALARM WAS NOTED ON THE INSULIN PUMP. THERE WAS INTERMITTENT BUTTON RESPONSE DUE TO MOISTURE DAMAGE TO THE KEYPAD TRACE. THE DEVICE WAS RECEIVED WITH SCRATCHED LCD WINDOW, CRACKED DISPLAY WINDOW CORNERS, AND A CRACKED RESERVOIR TUBE LIP.
IT WAS REPORTED THAT THE INSULIN PUMP ALARMED BUTTON ERROR AND BATTERY OUT LIMIT. THE CUSTOMER'S BLOOD GLUCOSE WAS 125 MG/DL. THE CUSTOMER STATED THAT THEY ALWAYS KEEP THE INSULIN PUMP IN THEIR BRA, AND NEVER HAD AN ISSUE WITH IT. NOTHING HAPPENED PRIOR TO THE BUTTON ERROR ALARM. THE CUSTOMER WAS ADVISED TO PLACE THE INSULIN PUMP IN A BABY SOCK IN ORDER TO AVOID MOISTURE EXPOSURE WHILE THE DEVICE IS BEING KEPT IN THE BRA. THE DISCONTINUATION OF THE DEVICE WAS ADVISED. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 447081 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-523NAP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 62 YR |