FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3972001 · Received July 31, 2014

Report

Report Number
3004209178-2014-88614
Event Type
Malfunction
Date Received
July 31, 2014
Date of Event
July 2, 2014
Report Date
July 2, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

NO BUTTON ERROR ALARM WAS NOTED ON THE INSULIN PUMP. THERE WAS INTERMITTENT BUTTON RESPONSE DUE TO MOISTURE DAMAGE TO THE KEYPAD TRACE. THE DEVICE WAS RECEIVED WITH SCRATCHED LCD WINDOW, CRACKED DISPLAY WINDOW CORNERS, AND A CRACKED RESERVOIR TUBE LIP.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED BUTTON ERROR AND BATTERY OUT LIMIT. THE CUSTOMER'S BLOOD GLUCOSE WAS 125 MG/DL. THE CUSTOMER STATED THAT THEY ALWAYS KEEP THE INSULIN PUMP IN THEIR BRA, AND NEVER HAD AN ISSUE WITH IT. NOTHING HAPPENED PRIOR TO THE BUTTON ERROR ALARM. THE CUSTOMER WAS ADVISED TO PLACE THE INSULIN PUMP IN A BABY SOCK IN ORDER TO AVOID MOISTURE EXPOSURE WHILE THE DEVICE IS BEING KEPT IN THE BRA. THE DISCONTINUATION OF THE DEVICE WAS ADVISED. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
447081 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-523NAP

Patients

Seq Age Sex Outcome Treatment
1 62 YR