PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 3004209178-2014-88606
- Event Type
- Malfunction
- Date Received
- July 31, 2014
- Date of Event
- July 2, 2014
- Report Date
- July 2, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI
- Reporter Occupation
- OTHER
Narratives
ALL OPERATING CURRENTS ARE WITHIN SPECIFICATION. NO UNEXPECTED LOW BATTERY, OFF NO POWER, OR BATTERY OUT LIMIT ALARMS NOTED. UNIT PASSED DISPLACEMENT TEST, REWIND, BASIC OCCLUSION, OCCLUSION, AND EXCESSIVE NO DELIVERY ALARM TEST. NO MOTOR ERROR ALARM NOTED. MOTOR PASSED MOTOR TEST. UNABLE TO PRIME DURING PRIME/A33 TEST DUE TO FAULTY FSR (GOLD). NO UNEXPECTED AUTO OFF ALARM NOTED DURING TESTING. UNIT RECEIVED WITH SCRATCHED LCD WINDOW, CRACKED RESERVOIR TUBE LIP, CRACKED BELT CLIP SLOT, AND CRACKED BATTERY TUBE THREADS.
CUSTOMER REPORTED A BLANK DISPLAY ON THE INSULIN PUMP. CUSTOMER ALSO STATED THAT THEY RECEIVED SEVERAL ERROR ALARMS INCLUDING A MOTOR ERROR ALARM DURING BASAL. CUSTOMER'S CURRENT BLOOD GLUCOSE READING IS 331 MG/DL. NO FURTHER INFORMATION REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 447080 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 67 YR |