FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3971998 · Received July 31, 2014

Report

Report Number
3004209178-2014-88606
Event Type
Malfunction
Date Received
July 31, 2014
Date of Event
July 2, 2014
Report Date
July 2, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ALL OPERATING CURRENTS ARE WITHIN SPECIFICATION. NO UNEXPECTED LOW BATTERY, OFF NO POWER, OR BATTERY OUT LIMIT ALARMS NOTED. UNIT PASSED DISPLACEMENT TEST, REWIND, BASIC OCCLUSION, OCCLUSION, AND EXCESSIVE NO DELIVERY ALARM TEST. NO MOTOR ERROR ALARM NOTED. MOTOR PASSED MOTOR TEST. UNABLE TO PRIME DURING PRIME/A33 TEST DUE TO FAULTY FSR (GOLD). NO UNEXPECTED AUTO OFF ALARM NOTED DURING TESTING. UNIT RECEIVED WITH SCRATCHED LCD WINDOW, CRACKED RESERVOIR TUBE LIP, CRACKED BELT CLIP SLOT, AND CRACKED BATTERY TUBE THREADS.

Description of Event or Problem · 1

CUSTOMER REPORTED A BLANK DISPLAY ON THE INSULIN PUMP. CUSTOMER ALSO STATED THAT THEY RECEIVED SEVERAL ERROR ALARMS INCLUDING A MOTOR ERROR ALARM DURING BASAL. CUSTOMER'S CURRENT BLOOD GLUCOSE READING IS 331 MG/DL. NO FURTHER INFORMATION REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
447080 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAS

Patients

Seq Age Sex Outcome Treatment
1 67 YR