PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 3004209178-2014-88577
- Event Type
- Malfunction
- Date Received
- July 31, 2014
- Date of Event
- July 2, 2014
- Report Date
- July 2, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
A CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS AND MINOR SCRATCHES ON THE DISPLAY WINDOW WERE NOTED DURING THE VISUAL INSPECTION. ALL BUTTONS FUNCTIONED PROPERLY. HOWEVER, MOISTURE DAMAGE WAS NOTED ON THE KEYPAD TRACES. NO MOISTURE DAMAGE WAS NOTED ON THE ELECTRONIC ASSEMBLY. NO BLACK DOTS WERE NOTED ON THE DISPLAY.
IT WAS REPORTED THAT THE CUSTOMER'S INSULIN PUMP WAS EXPOSED TO WATER AND THEN THE CUSTOMER WAS UNABLE TO NAVIGATE THE MENUS ON THE SCREEN, AS A BLACK DOT HAD APPEARED. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE AND TO REVERE TO A BACK-UP PLAN. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 447226 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-754CMS |