FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3971958 · Received July 31, 2014

Report

Report Number
3004209178-2014-88583
Event Type
Malfunction
Date Received
July 31, 2014
Date of Event
May 10, 2014
Report Date
May 10, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Report Source
Manufacturer report
Reporter Location
NY
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

NO BUTTON ERROR ALARM. INTERMITTENT BUTTON RESPONSE DUE TO MOISTURE DAMAGE KEYPAD TRACE. SCRATCHED LCD WINDOW, CRACKED DISPLAY WINDOW CORNERS, BATTERY TUBE THREADS CRACKED, CRACKED RESERVOIR TUBE LIP, RESERVOIR TUBE CRACKED.

Description of Event or Problem · 1

CUSTOMER REPORTED ISSUES WITH THE INSULIN PUMP. CUSTOMER WAS GETTING THE BUTTON ERROR ALARM. THE BLOOD GLUCOSE READING IS 217 MG/DL. THE INSULIN PUMP WILL BE REPLACED. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
446899 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-523NAL

Patients

Seq Age Sex Outcome Treatment
1 53 YR