FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3971923 · Received July 31, 2014

Report

Report Number
3004209178-2014-88591
Event Type
Malfunction
Date Received
July 31, 2014
Date of Event
July 2, 2014
Report Date
July 2, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A33 ALARM DURING PRIME/A33 TEST AT 4.0 LBS DUE TO FAULTY FSR (GOLD) AND MOTOR ERROR ALARM DURING BASIC OCCLUSION TEST DUE TO FAULTY FSR (GOLD). MOTOR PASSED MOTOR TEST.PUMP RECEIVED WITH SCRATCHED LCD WINDOW, CRACKED CASE DISPLAY WINDOW CORNER, CRACKED BATTERY TUBE THREADS, CRACKED RESERVOIR TUBE LIP, CRACKED RESERVOIR TUBE, CRACKED RESERVOIR TUBE WINDOW AND A MISSING END CAP STICKER.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED DURING THE MANUAL PRIME. CUSTOMER ALSO REPORTED A MOTOR ERROR ALARM DURING BASAL. CUSTOMER STATED THAT THE INSULIN PUMP WAS DROPPED. TROUBLESHOOTING WAS PERFORMED AND THE DRIVE SUPPORT CAP APPEARED TO BE NORMAL. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
446953 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-523NAL

Patients

Seq Age Sex Outcome Treatment
1 36 YR