PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 3004209178-2014-88590
- Event Type
- Malfunction
- Date Received
- July 31, 2014
- Date of Event
- May 16, 2014
- Report Date
- May 16, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UT
- Reporter Occupation
- PATIENT
Narratives
NO A17 ALARM OR A21 ALARM NOTED. UNIT PASSED THE SELF TEST AND A21 ERROR TEST. PUMP RECEIVED WITH A47 ALARM IN THE HISTORY FILE. A47 ALARM DUE TO CORRUPTED HISTORY FILE. PUMP PASSED THE REWIND, BASIC OCCLUSION, PRIME/A33 AND DISPLACEMENT TESTS. PUMP RECEIVED WITH STUCK MOTOR ERROR ALARM LOOP DURING BOLUS DELIVERY. THE MOTOR WAS TESTED OUTSIDE OF THE DEVICE AND PASSED. THE MOTOR MAY HAVE HAD AN INTERMITTENT FAILURE THAT WAS NOT DETECTED DURING OUR TESTING. UNABLE TO CONFIRM E70 ALARM DUE TO ERASED HISTORY FILE. UNIT HAD SCRATCHED DISPLAY WINDOW AND CRACKED DISPLAY WINDOW CORNER.
CUSTOMER REPORTED ISSUES WITH THE INSULIN PUMP. CUSTOMER STATED THAT THERE IS A MOTOR ERROR ALARM. ASSISTED CUSTOMER IN CLEARING THE ALARM BY DISCONNECTING FROM THE INSULIN PUMP. FURTHERMORE, CUSTOMER RECEIVED VARIOUS OTHER ALARMS. CUSTOMER'S BLOOD GLUCOSE IS NOT PROVIDED. NOTHING FURTHER REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 446429 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-722LNAH |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 29 YR |