FDA Adverse Event
Malfunction
Summary report: N
530G INSULIN PUMP
MDR report key: 3971921
·
Received July 31, 2014
Report
- Report Number
- 3004209178-2014-88580
- Event Type
- Malfunction
- Date Received
- July 31, 2014
- Date of Event
- July 2, 2014
- Report Date
- July 2, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZO
- PMA / PMN Number
- 120010
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
UNIT PASSED REWIND, BASIC OCCLUSION TEST, OCCLUSION TEST, PRIME/A33TEST, EXCESSIVE NO DELIVERY TEST AND DISPLACEMENT TEST. UNIT RECEIVED WITH INTERMITTENT ACT BUTTON DUE TO FLATTENED DOME SWITCH. (CREASED) NO BUTTON ERROR ALARMS NOTED. UNIT RECEIVED WITH SCRATCHED LCD WINDOW. (B)(4).
Description of Event or Problem · 1
CUSTOMER STATED THAT BOLUS DOES NOT WORK, AND THEY HAVE TO CONSTANTLY PUSH THE BOLUS BUTTON. THE BLOOD GLUCOSE READING IS 134 MG/DL. THE INSULIN PUMP WILL BE REPLACED. NOTHING FURTHER REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 446438 | 530G INSULIN PUMP | OZO | OZO | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-751NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 26 YR |