FDA Adverse Event Malfunction Summary report: N

530G INSULIN PUMP

MDR report key: 3971921 · Received July 31, 2014

Report

Report Number
3004209178-2014-88580
Event Type
Malfunction
Date Received
July 31, 2014
Date of Event
July 2, 2014
Report Date
July 2, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZO
PMA / PMN Number
120010
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

UNIT PASSED REWIND, BASIC OCCLUSION TEST, OCCLUSION TEST, PRIME/A33TEST, EXCESSIVE NO DELIVERY TEST AND DISPLACEMENT TEST. UNIT RECEIVED WITH INTERMITTENT ACT BUTTON DUE TO FLATTENED DOME SWITCH. (CREASED) NO BUTTON ERROR ALARMS NOTED. UNIT RECEIVED WITH SCRATCHED LCD WINDOW. (B)(4).

Description of Event or Problem · 1

CUSTOMER STATED THAT BOLUS DOES NOT WORK, AND THEY HAVE TO CONSTANTLY PUSH THE BOLUS BUTTON. THE BLOOD GLUCOSE READING IS 134 MG/DL. THE INSULIN PUMP WILL BE REPLACED. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
446438 530G INSULIN PUMP OZO OZO MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-751NAS

Patients

Seq Age Sex Outcome Treatment
1 26 YR