PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 3971920
·
Received July 31, 2014
Report
- Report Number
- 3004209178-2014-88588
- Event Type
- Malfunction
- Date Received
- July 31, 2014
- Date of Event
- July 2, 2014
- Report Date
- July 2, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
FROZEN SCREEN DUE TO MOISTURE DAMAGE ON THE ELECTRONICS. MOISTURE DAMAGE FOUND ON THE MOTOR ALSO. UNABLE TO VERIFY THE ACTIVATION ANOMALY DUE TO FROZEN SCREEN. NO HIGH BEEP OR AUDIO BEEP ANOMALY NOTED. PUMP RECEIVED WITH CRACKED DISPLAY WINDOW CORNER, RESERVOIR TUBE LIP, SCRATCHED LCD WINDOW.
Description of Event or Problem · 1
CUSTOMER REPORTED THAT WHEN INSERTING A NEW BATTERY, THE PUMP GIVES A HIGH BEEP. THE BLOOD GLUCOSE READING IS 18 MG/DL. NOTHING FURTHER REPORTED.