FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3971920 · Received July 31, 2014

Report

Report Number
3004209178-2014-88588
Event Type
Malfunction
Date Received
July 31, 2014
Date of Event
July 2, 2014
Report Date
July 2, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FROZEN SCREEN DUE TO MOISTURE DAMAGE ON THE ELECTRONICS. MOISTURE DAMAGE FOUND ON THE MOTOR ALSO. UNABLE TO VERIFY THE ACTIVATION ANOMALY DUE TO FROZEN SCREEN. NO HIGH BEEP OR AUDIO BEEP ANOMALY NOTED. PUMP RECEIVED WITH CRACKED DISPLAY WINDOW CORNER, RESERVOIR TUBE LIP, SCRATCHED LCD WINDOW.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT WHEN INSERTING A NEW BATTERY, THE PUMP GIVES A HIGH BEEP. THE BLOOD GLUCOSE READING IS 18 MG/DL. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
446952 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-754LWWL

Patients

Seq Age Sex Outcome Treatment
1