OSCILLATING SAW ATTACHMENT II
Report
- Report Number
- 8030965-2014-10440
- Event Type
- Malfunction
- Date Received
- July 31, 2014
- Report Date
- July 3, 2014
- Manufacturer
- SYNTHES OBERDORF
- Product Code
- GFA
- PMA / PMN Number
- PEXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
EVENT DATE: UNKNOWN. DEVICE IS AN INSTRUMENT AND IS NOT IMPLANTED/EXPLANTED. (B)(6). THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
A SERVICE HISTORY REVIEW WAS CONDUCTED. THE REPORT INDICATES THAT THE DEVICE HAS NOT BEEN SERVICED DURING THE PAST 6 MONTHS. THERE IS NO INFORMATION RELEVANT TO THE CURRENT COMPLAINT ISSUE. NO HISTORY COULD BE FOUND. THE INVESTIGATION IS BASED ON THE SERVICE & REPAIR RECORD RECEIVED. THE DEVICE FAILURE REPORTED BY THE CUSTOMER COULD BE CONFIRMED. THE SERVICE HISTORY REVIEW SHOWS NO PREVIOUS SERVICE CONDITIONS THAT ARE RELEVANT TO THE CURRENT COMPLAINT ISSUE. THE SERVICE TECHNICIAN NOTED THE FOLLOWING ACTION TAKEN REPAIR. DURING THE PRE-REPAIR DIAGNOSTIC ASSESSMENT THE SERVICE TECHNICIAN IDENTIFIED THE FOLLOWING FAILURE: BEARING NOT FUNCTIONING, DEFECTIVE. THE ROOT CAUSE OF THE EVENT WAS DETERMINED TO BE NORMAL WEAR. THE DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
DEVICE REPORT FROM SYNTHES (B)(4) REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT THE ATTACHMENT DOES NOT FUNCTION. THERE IS AN ISSUE WITH BEARINGS OF SAW ATTACHMENT AND POSSIBLY AN ISSUE WITH ENGINE. THIS IS REPORT 1 OF 1 FOR (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 447863 | OSCILLATING SAW ATTACHMENT II | BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL | GFA | SYNTHES OBERDORF |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |