FDA Adverse Event Injury Summary report: N

DXTEND SCREW LOCK D4.5X30MM

MDR report key: 3971687 · Received July 31, 2014

Report

Report Number
1818910-2014-24713
Event Type
Injury
Date Received
July 31, 2014
Date of Event
July 14, 2014
Report Date
July 14, 2014
Manufacturer
DEPUY FRANCE S.A.S.-SAINT PRIEST 3003895575
Product Code
KWS
PMA / PMN Number
PK120174
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

THE DEVICES ASSOCIATED WITH THIS REPORT WERE NOT RETURNED. THE DHR ANALYSIS OF THE BATCHES PROVIDED SHOWS AN INITIAL CONFORMANCE OF THESE PRODUCTS WITH REGARDS TO THEIR SPECIFICATION. FOR THESE BATCHES, THERE WAS NO DEVIATION OR NON-CONFORMANCE. REQUESTS FOR ADDITIONAL INVESTIGATIONAL INPUTS WERE MADE IN ACCORDANCE WITH WI-7915 APPENDIX A. NO ADDITIONAL INFORMATION WAS OBTAINED. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY PRODUCT CONTRIBUTION TO THE REPORTED EVENT WITH THE INFORMATION PROVIDED. BASED ON THE INABILITY TO IDENTIFY ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.

Description of Event or Problem · 1

REVISION: TOTAL REVERSE SHOULDER, PERFORMED BY (B)(6) ON (B)(6) 2014 AT (B)(6). REASON: INSTABILITY, INCORRECT VERSION, PAIN AND EMOTIONAL STRESS. PRIMARY: DELTA XTEND, PERFORMED BY (B)(6).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
447814 DXTEND SCREW LOCK D4.5X30MM SHOULDER OTHER IMPLANT KWS DEPUY FRANCE S.A.S.-SAINT PRIEST 3003895575 5124376

Patients

Seq Age Sex Outcome Treatment
1 53 YR Required Intervention