HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Report
- Report Number
- 1416980-2014-24900
- Event Type
- Malfunction
- Date Received
- July 31, 2014
- Date of Event
- July 7, 2014
- Report Date
- July 8, 2014
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). AS THE CASSETTE WAS NOT RETURNED AND THE LOT NUMBER WAS UNKNOWN, A DEVICE ANALYSIS COULD NOT BE COMPLETED. HOWEVER, THE HOME PATIENT (HP) REPORTED THAT A SUPPLY BAG FELL AND DISCONNECTED WHICH IS A KNOWN CAUSE OF THE ALARM. USE ERROR AND PROPER USER INSTRUCTIONS ARE ADDRESSED IN ¿THE HOMECHOICE AND HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE¿ WHICH IS SHIPPED WITH EVERY HOMECHOICE DEVICE. IT INSTRUCTS THE USER TO PLACE SOLUTION BAGS ON A FLAT STABLE SURFACE. FURTHERMORE, IT WARNS THAT FALLING BAGS CAN RESULT IN A DISCONNECTION WHICH CAN LEAD TO CONTAMINATION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT A SYSTEM ERROR (SE) 2240 ALARM (AIR IN LINE) OCCURRED ON A HOMECHOICE (HC) DEVICE DURING DWELL 3 OF 5 OF AUTOMATED PERITONEAL DIALYSIS (APD) THERAPY. THE HOME PATIENT (HP) WAS CONNECTED AT THE TIME OF THE ALARM. TROUBLESHOOTING REVEALED THAT THE SUPPLY BAG ON THE LAST FILL LINE FELL AND DISCONNECTED. A TECHNICAL SERVICE REPRESENTATIVE (TSR) REVIEWED PROPER PROCEDURES WITH THE HP. THE HP DID NOT WANT TO RESTART THERAPY AND PLANNED ON CONTACTING A REGISTERED NURSE (RN) REGARDING THE ISSUE. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 448141 | HOMECHOICE AUTOMATED PD SET WITH CASSETTE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 79 YR | HOMECHOICE |