FDA Adverse Event Injury Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3971640 · Received July 31, 2014

Report

Report Number
2531779-2014-22003
Event Type
Injury
Date Received
July 31, 2014
Report Date
July 17, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

PRODUCT ANALYSIS: THE DEVICE WAS RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 09/02/2014 WITH THE FOLLOWING FINDINGS: THE COMPLAINT COULD NOT BE DUPLICATED OR CONFIRMED WITH INVESTIGATION. REVIEW OF THE PUMP¿S BLACK BOX FOUND THE LAST BOLUS DELIVERY OCCURRED ON (B)(6) 2014; THERE WAS A LOSS OF PRIME EVENT ON (B)(6) 204 AND THE PUMP RESUMED INSULIN DELIVERY ON (B)(6) 2014; THE LAST BASAL DELIVERY OCCURRED ON (B)(6) 2014. THE PUMP SUCCESSFULLY COMPLETED A PRIME SEQUENCE, BOLUS DELIVERIES, DELIVERY ACCURACY TEST, AND 24-HOUR EXERCISE TEST WITHOUT ISSUE OR ALARM. BASAL AND BOLUS DELIVERIES DURING INVESTIGATION WERE ACCURATELY RECORDED IN THE PUMP¿S TOTAL DAILY DOSE HISTORY. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Additional Manufacturer Narrative · 1

THE DEVICE HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2014, IT WAS REPORTED THE PATIENT¿S BLOOD GLUCOSE WAS 359 MG/DL WITH MODERATE KETONE LEVELS ON AN UNSPECIFIED DATE. IT WAS REPORTED THERE WAS AN INACCURATE DELIVERY ISSUE WITH THE PUMP. NO ADDITIONAL INFORMATION WAS PROVIDED AND TROUBLESHOOTING COULD NOT BE COMPLETED. THIS COMPLAINT IS BEING REPORTED BECAUSE THE PATIENT¿S HYPERGLYCEMIA WAS REPORTEDLY ASSOCIATED WITH AN ALLEGED PUMP MALFUNCTION OF UNKNOWN CAUSE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
447713 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 49 YR Life Threatening