FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 3971627 · Received July 31, 2014

Report

Report Number
1416980-2014-24890
Event Type
Malfunction
Date Received
July 31, 2014
Date of Event
July 8, 2014
Report Date
July 8, 2014
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS NOT RETURNED AND THE LOT NUMBER WAS UNKNOWN; THEREFORE, A DEVICE ANALYSIS COULD NOT BE COMPLETED. HOWEVER, THE PATIENT CONNECTING TO AN INCOMPLETE PRIME IS A KNOWN CAUSE OF THIS ALARM. THE HOMECHOICE AND HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE, WHICH IS SHIPPED WITH EVERY HOMECHOICE DEVICE, PROVIDES STEP-BY-STEP INSTRUCTIONS FOR PROPERLY PRIMING THE DISPOSABLE SET. IT WARNS THE USER NOT TO CONNECT TO THE PATIENT LINE UNLESS THE FLUID LEVEL IS AT OR NEAR THE CONNECTOR AT THE END OF THE DISPOSABLE SET PATIENT LINE. A REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A SYSTEM ERROR (SE) 2240 ALARM (AIR IN SET / LINE) OCCURRED ON THE HOMECHOICE (HC) DEVICE DURING THE INITIAL DRAIN. THE HOME PATIENT (HP) WAS CONNECTED AT THE TIME OF THE ALARM. DURING THE TROUBLESHOOTING, THE HP INFORMED THE TECHNICAL SERVICE REPRESENTATIVE (TSR) THAT THEY CONNECTED PRIOR TO PRIMING BEING COMPLETED. THE TSR EXPLAINED THE ALARM AND INSTRUCTED THE HP TO CYCLE THE POWER ON THE HC TO CLEAR THE ALARM. THE TSR ADVISED THE HP TO START OVER WITH NEW SUPPLIES. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
447584 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 89 YR HOMECHOICE