FDA Adverse Event Malfunction Summary report: N

G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM

MDR report key: 3971455 · Received July 31, 2014

Report

Report Number
3004753838-2014-25121
Event Type
Malfunction
Date Received
July 31, 2014
Date of Event
July 3, 2014
Report Date
July 7, 2014
Manufacturer
DEXCOM, INC.
Product Code
MDS
PMA / PMN Number
P120005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION OF THE RETURNED DEVICE CONFIRMED THE REPORTED NO VIBRATION. THE ROOT CAUSE WAS DETERMINED TO BE A DEFECTIVE VIBRATE MOTOR DUE TO HIGH VIBRATOR LOAD RESISTANCE.

Description of Event or Problem · 1

PATIENT CONTACTED DEXCOM TECHNICAL SUPPORT ON (B)(6) 2014 TO CLAIM NO VIBRATION ON (B)(6) 2014. AT THE TIME OF THE CALL TO DEXCOM TECHNICAL SUPPORT, NO MEDICAL INTERVENTION OR INJURIES WERE REPORTED. ADDITIONALLY, DURING THE CALL, DEXCOM TECHNICAL SUPPORT ADVISED PATIENT TO TEST THE ALERT FUNCTIONALITY AND REPORTED ALERTS WERE NOT FUNCTIONAL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
447127 G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM MDS MDS DEXCOM, INC. MT20649

Patients

Seq Age Sex Outcome Treatment
1 19 YR