HEARTWARE VENTRICULAR ASSIST SYSTEM
Report
- Report Number
- 3007042319-2014-00746
- Event Type
- Malfunction
- Date Received
- July 31, 2014
- Date of Event
- June 16, 2014
- Report Date
- July 3, 2014
- Manufacturer
- HEARTWARE
- Product Code
- DSQ
- PMA / PMN Number
- P100047
- Removal / Correction Number
- Z-1607-2014
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UT, US
- Reporter Occupation
- OTHER
Narratives
THE HEARTWARE VENTRICULAR ASSIST DEVICE (VAD) IS USED FOR TREATMENT NOT DIAGNOSIS. IT WAS REPORTED THAT THE PATIENT'S BATTERIES DISCONNECTED SUDDENLY. THERE WAS NO REPORTED PATIENT INJURY RELATED TO THIS EVENT. TWO (2) BATTERIES WERE RETURNED TO THE MANUFACTURER WHILE THE THIRD WAS NOT RETURNED. THE BATTERY ((B)(4)) WAS REMOVED FROM THE MARKET AS PART OF FSCA APR2014.1 AND WAS DESTROYED AS PART OF THE REQUIRED ACTIONS. AS A RESULT, THE DEVICE IS NOT AVAILABLE FOR ANALYSIS. BASED ON THE RESULTS OF THE PREVIOUS MANUFACTURER'S INTERNAL INVESTIGATION, FIELD ACTIONS AND CHANGES TO THE BATTERY INTERNAL CELL SUPPLIER, NO ADDITIONAL INVESTIGATION OF THIS EVENT IS REQUIRED AT THIS TIME. THE MOST LIKELY CAUSE OF THE REPORTED EVENT CAN BE ATTRIBUTED TO FAULTY INTERNAL CELLS WITHIN THE BATTERY PACK. Z-1607-2014. FIELD SAFETY NOTICE (FSCA APR2014) WAS ISSUED TO PROVIDE PATIENTS AND HEALTHCARE PROVIDERS WITH INFORMATION TO RECOGNIZE BATTERIES WITH LESS THAN TWO HOURS OF RUN TIME, AS WELL AS REEMPHASIZE INSTRUCTION ON ACTIONS TO TAKE WHEN BATTERY ALARMS OCCUR AND REINFORCE PROPER POWER MANAGEMENT FIELD SAFETY NOTICE WAS THEN EXPANDED (FSCA APR2014.1) IN ORDER TO REMOVE BATTERIES FROM THE FIELD THAT WERE RELEASED PRIOR TO THE IMPLEMENTATION OF ENHANCED BATTERY SCREENING PROCESS TO ADDRESS AND PREVENT BATTERY FAILURES. THE VENTRICULAR ASSIST SYSTEM IS INDICATED FOR USE AS A BRIDGE TO CARDIAC TRANSPLANTATION IN PATIENTS WHO ARE AT RISK OF DEATH FROM REFRACTORY END-STAGE LEFT VENTRICULAR HEART FAILURE. THE SYSTEM IS DESIGNED FOR IN-HOSPITAL AND OUT-OF-HOSPITAL SETTINGS, INCLUDING TRANSPORTATION. THE INSTRUCTIONS FOR USE (IFU), PATIENT MANUAL, AND TRAINING MATERIAL PROVIDE CLEAR INSTRUCTIONS TO THE USER ON PROPER USAGE AND CARE OF POWER SOURCES. THE IFU PROVIDES INSTRUCTION TO FURTHER EDUCATE THE PATIENT ABOUT PRODUCT SAFETY, VISUAL AND AUDIBLE ALARM MANAGEMENT, AND DEVICE MANAGEMENT; ADDITIONAL GUIDELINES INSTRUCT THE USER ON HOW TO DETECT AND REACT TO A POWER SOURCE MALFUNCTION. ADDITIONALLY THERE IS A WARNING TO KEEP SPARE, FULLY CHARGED BATTERIES AND BACK UP CONTROLLER AVAILABLE AT ALL TIME. THE STEPS FOR EXCHANGE OF BATTERIES AND CONTROLLERS ARE OUTLINED. HEARTWARE IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO FDA WARNING LETTER FLA-14-14, DATED JUNE 2, 2014, AND PURSUANT TO THE PROVISIONS OF 21 CFR PART 803. THIS IS ONE OF TWO REPORTS (3007042319-2014-00746 AND 3007042319-2014-00899) SUBMITTED FOR DEVICES RELATED TO THE SAME EVENT.
THE DEVICE IS AVAILABLE FOR EVALUATION, BUT HAS NOT BEEN RECEIVED BY THE MANUFACTURER. ADDITIONAL INFORMATION WILL BE SUBMITTED WITHIN THIRTY (30) DAYS OF RECEIPT. PRODUCT IS IN ROUTE.
APPROXIMATELY EIGHTEEN MONTHS AFTER IMPLANT, THIS PATIENT EXPERIENCED A LOSS OF POWER WHEN THE BATTERIES DISCONNECTED SUDDENLY. FROM THE REPORT, IT WAS NOT CLEAR WHETHER THIS WAS A PHYSICAL DISCONNECTION OR AN ELECTRICAL DISCONNECTION. PRELIMINARY LOG FILE ANALYSIS CONFIRMED AN ASSOCIATED CONTROLLER SHUT-DOWN AND PUMP STOP. THE BATTERIES INVOLVED WERE EXCHANGED WITHOUT ANY REPORTED INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 447240 | HEARTWARE VENTRICULAR ASSIST SYSTEM | CIRCULATORY ASSIST SYSTEM, BATTERY | DSQ | HEARTWARE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | (B)(4) - BATTERY |