EASYTRAK 2
MDR report key: 3971329
·
Received July 31, 2014
Report
- Report Number
- 2124215-2014-10652
- Event Type
- Injury
- Date Received
- July 31, 2014
- Date of Event
- March 14, 2014
- Report Date
- April 16, 2014
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- NIK
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
(B)(4). AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.
Description of Event or Problem · 1
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS LEFT VENTRICULAR (LV) LEAD EXHIBITED HIGH OUT-OF-RANGE PACING LEAD IMPEDANCE, LOSS OF CAPTURE (LOC) INTERMITTENT SENSING AFTER THE PATIENT FELL DUE TO BICYCLE ACCIDENT. THE LEAD HAD GREATER THAN 2000 OHMS IMPEDANCE WHEN TESTED WITH PACING SYSTEM ANALYZER (PSA) AND APPEARED TO BE FRACTURED. THE LEAD WAS CAPPED AND REPLACED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 446374 | EASYTRAK 2 | IMPLANTABLE LEAD | NIK | CPI - DEL CARIBE | 4543 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 61 YR | Hospitalization| L| R | 4470| 4543| 0185| N119 |