FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 3971162 · Received April 29, 2014

Report

Report Number
8020893-2014-01017
Event Type
Malfunction
Date Received
April 29, 2014
Report Date
April 1, 2014
Manufacturer
COVIDIEN, FORMERLY NELLCOR
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NO INFORMATION

Narratives

Additional Manufacturer Narrative · 1

THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) INSPECTED THE DEVICE AND VERIFIED THE MALFUNCTION. THE CSE REPLACED THE GRAPHIC USER INTERFACE (GUI) CENTRAL PROCESSING UNIT (CPU) PRINTED CIRCUIT BOARD (PCB) AND UPGRADED THE SOFTWARE TO THE CURRENT REVISION. THE CSE PERFORMED EXTENDED SELF-TESTING ON THE DEVICE AND ALL TESTS PASSED. (B)(4).

Description of Event or Problem · 1

A REPORT WAS RECEIVED FROM THE USA WITH INFORMATION STATING THAT WHEN AN 840 VENTILATOR WAS POWERED ON THE SCREEN WAS BLANK. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME THE MALFUNCTION OCCURRED. THE DATE OF THE EVENT WAS NOT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
256480 840 VENTILATOR CBK COVIDIEN, FORMERLY NELLCOR 840

Patients

Seq Age Sex Outcome Treatment
1