MINIMAL ACCESS ATTACHMENT
Report
- Report Number
- 1045834-2014-12536
- Event Type
- Malfunction
- Date Received
- July 31, 2014
- Date of Event
- July 18, 2014
- Report Date
- July 18, 2014
- Manufacturer
- DEPUY SYNTHES POWER TOOLS
- Product Code
- HBE
- PMA / PMN Number
- PK042783
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
ADDITIONAL NARRATIVE:THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE AND OBSERVED THAT ATTACHEMENT HAD A STUCK SLEEVE LOCKING MECHANISM. THEREFORE, THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO NORMAL WEAR AND SERVICING OVER TIME. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
IT WAS REPORTED THAT DURING CRANIOTOMY SURGERY, IT WAS OBSERVED THAT THE ATTACHMENT DEVICE WAS STICKING. THERE WAS A TWENTY MINUTE DELAY TO THE SURGICAL PROCEDURE DUE TO THE TIME TO RETRIEVE AND SET-UP AN IDENTICAL SPARE DEVICE. THE SURGERY WAS SUCCESSFULLY COMPLETED WITH THE SPARE DEVICE. THERE WAS PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 448196 | MINIMAL ACCESS ATTACHMENT | DRILLS, BURRS, TREPHINES & ACCESSORIES - ATTACHMENT | HBE | DEPUY SYNTHES POWER TOOLS | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |