FDA Adverse Event Malfunction Summary report: N

MINIMAL ACCESS ATTACHMENT

MDR report key: 3970980 · Received July 31, 2014

Report

Report Number
1045834-2014-12536
Event Type
Malfunction
Date Received
July 31, 2014
Date of Event
July 18, 2014
Report Date
July 18, 2014
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HBE
PMA / PMN Number
PK042783
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL NARRATIVE:THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE AND OBSERVED THAT ATTACHEMENT HAD A STUCK SLEEVE LOCKING MECHANISM. THEREFORE, THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO NORMAL WEAR AND SERVICING OVER TIME. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING CRANIOTOMY SURGERY, IT WAS OBSERVED THAT THE ATTACHMENT DEVICE WAS STICKING. THERE WAS A TWENTY MINUTE DELAY TO THE SURGICAL PROCEDURE DUE TO THE TIME TO RETRIEVE AND SET-UP AN IDENTICAL SPARE DEVICE. THE SURGERY WAS SUCCESSFULLY COMPLETED WITH THE SPARE DEVICE. THERE WAS PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
448196 MINIMAL ACCESS ATTACHMENT DRILLS, BURRS, TREPHINES & ACCESSORIES - ATTACHMENT HBE DEPUY SYNTHES POWER TOOLS NA

Patients

Seq Age Sex Outcome Treatment
1