FDA Adverse Event Malfunction Summary report: N

ACCU-CHEK ® COMPACT TEST DRUM

MDR report key: 3970943 · Received July 31, 2014

Report

Report Number
1823260-2014-05695
Event Type
Malfunction
Date Received
July 31, 2014
Date of Event
July 6, 2014
Report Date
September 10, 2014
Manufacturer
ROCHE DIAGNOSTICS
Product Code
LFR
PMA / PMN Number
NA
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THE EVENT OCCURRED IN (B)(6).

Description of Event or Problem · 1

CUSTOMER REPORTEDLY RECEIVED RESULTS OF 0.9 MMOL/L AND 5.6 MMOL/L WITHIN 10 MINUTES ON THE COMPACT PLUS SYSTEM. NO ACTIONS TAKEN BASED ON DEVICE RESULTS. NO ADVERSE EVENT REPORTED. REQUESTED RETURN OF SUSPECT DEVICE AND REPLACEMENT WAS SENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
447961 ACCU-CHEK ® COMPACT TEST DRUM BLOOD GLUCOSE MONITORING TEST STRIPS LFR ROCHE DIAGNOSTICS NA 20808943

Patients

Seq Age Sex Outcome Treatment
1 074 YR