LC PCA SW VERSION 1

MDR report key: 3970762 · Received April 29, 2014

Report

Report Number
9615050-2014-03067
Event Type
Malfunction
Date Received
April 29, 2014
Date of Event
March 18, 2014
Report Date
March 18, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
MEA
PMA / PMN Number
K042800
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT COMPLETE. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE DEVICE WAS RETURNED TO THE SERVICE CENTER FOR A REPORT FROM THE CUSTOMER CONTACT THAT THE DEVICE DID NOT PASS THE ELECTRICAL SAFETY TEST. THIS IS NOT A REPORTABLE MALFUNCTION. HOWEVER, DURING VERIFICATION TESTING AT THE SERVICE CENTER, IT WAS NOTED THAT THE GROUND PRONG ON THE AC POWER CORD WAS LOOSE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
256325 LC PCA SW VERSION 1 80MEA MEA HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA