FDA Adverse Event Injury Summary report: N

M2A-MAGNUM PF CUP 54ODX48ID

MDR report key: 3970549 · Received July 31, 2014

Report

Report Number
0001825034-2014-06639
Event Type
Injury
Date Received
July 31, 2014
Date of Event
October 22, 2012
Report Date
June 30, 2016
Manufacturer
BIOMET ORTHOPEDICS
Product Code
KWA
PMA / PMN Number
PK042037
Removal / Correction Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
ATTORNEY

Narratives

Additional Manufacturer Narrative · 1

CURRENT INFORMATION IS INSUFFICIENT TO PERMIT A CONCLUSION AS TO THE CAUSE OF THE EVENT. REVIEW OF DEVICE HISTORY RECORDS SHOW THAT LOT RELEASED WITH NO RECORDED ANOMALY OR DEVIATION. THERE ARE WARNINGS IN THE PACKAGE INSERT THAT STATE THAT THIS TYPE OF EVENT CAN OCCUR: UNDER POSSIBLE ADVERSE EFFECTS, NUMBER 1 STATES, ¿MATERIAL SENSITIVITY REACTIONS.¿ NUMBER 6 STATES, ¿INADEQUATE RANGE OF MOTION DUE TO IMPROPER SELECTION OR POSITIONING OF COMPONENTS.¿ NUMBER 14 STATES, ¿INTRAOPERATIVE OR POSTOPERATIVE BONE FRACTURE AND/OR POSTOPERATIVE PAIN.¿ NUMBER 15 STATES, ¿ELEVATED METAL ION LEVELS HAVE BEEN REPORTED WITH METAL-ON-METAL ARTICULATING SURFACES.¿ THIS REPORT IS BASED ON ALLEGATIONS SET FORTH IN PLAINTIFF¿S COMPLAINT AND THE ALLEGATIONS CONTAINED THEREIN ARE UNVERIFIED. THIS REPORT IS NUMBER 1 OF 2 MDRS FILED FOR THE SAME EVENT (REFERENCE 1825034-2014-06639 & 06640).

Additional Manufacturer Narrative · 1

THIS FOLLOW-UP REPORT IS BEING FILED TO RELAY ADDITIONAL INFORMATION, WHICH WAS UNKNOWN AT THE TIME OF THE INITIAL MEDWATCH.

Additional Manufacturer Narrative · 1

THIS FOLLOW-UP REPORT IS BEING FILED TO CORRECT AND RELAY ADDITIONAL INFORMATION, WHICH WAS UNKNOWN AT THE TIME OF THE INITIAL MEDWATCH. (B)(4). NOT RETURNED BY ATTORNEY.

Description of Event or Problem · 1

LEGAL COUNSEL FOR PATIENT REPORTED PATIENT UNDERWENT A LEFT TOTAL HIP ARTHROPLASTY ON (B)(6) 2008. PATIENT'S LEGAL COUNSEL FURTHER REPORTED PATIENT ALLEGATIONS OF PAIN, DAMAGE TO BONE/TISSUE, LACK OF MOBILITY, METAL POISONING, METALLOSIS AND ELEVATED METAL ION LEVELS. SUBSEQUENTLY, PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2012. A REVIEW OF THE INVOICE HISTORY CONFIRMED THE SURGERY DATES AND INDICATES THE MODULAR HEAD, TAPER ADAPTER AND ACETABULAR CUP WERE REMOVED AND REPLACED. THIS REPORT IS BASED ON ALLEGATIONS SET FORTH IN PLAINTIFF¿S COMPLAINT AND THE ALLEGATIONS CONTAINED THEREIN ARE UNVERIFIED.

Description of Event or Problem · 1

LEGAL COUNSEL FOR PATIENT REPORTED PATIENT UNDERWENT A LEFT TOTAL HIP ARTHROPLASTY ON (B)(6) 2008. PATIENT'S LEGAL COUNSEL FURTHER REPORTED PATIENT ALLEGATIONS OF PAIN, DAMAGE TO BONE/TISSUE, LACK OF MOBILITY, METAL POISONING, METALLOSIS AND ELEVATED METAL ION LEVELS. SUBSEQUENTLY, PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2012. A REVIEW OF THE INVOICE HISTORY CONFIRMED THE SURGERY DATES AND INDICATES THE MODULAR HEAD, TAPER ADAPTER AND ACETABULAR CUP WERE REMOVED AND REPLACED. THIS REPORT IS BASED ON ALLEGATIONS SET FORTH IN PLAINTIFF'S COMPLAINT AND THE ALLEGATIONS CONTAINED THEREIN ARE UNVERIFIED. ADDITIONAL INFORMATION RECEIVED IN PATIENT OPERATIVE (OP) NOTES REPORTS PATIENT WAS REVISED DUE TO METALLOSIS WITH EARLY LOOSENING OF ACETABULUM. REVISION OPERATIVE NOTES REPORT THE PRESENCE OF EPITHELIALIZED SAC OF BURSAL FLUID, VERTICAL CUP, ARTHROFIBROTIC TISSUE, EXTERNAL ROTATOR MASS ON THE FEMUR, THINLY VISCOUS GRAY FLUID, CHARCOAL COLORED TRUNNION, CAVITARY DEFECT, METALLOSIS, METAL STAINING, AND SCLEROTIC ACETABULUM. IN ADDITION, OP NOTES REPORT THE ORIGINAL REAMING FOR THE ACETABULAR COMPONENT HAD NOT COME OUT AND HAD NOT BEEN CARRIED OUT ALL THE WAY TO THE TRUE DEPTH OF THE ACETABULUM AND THAT THE HORSESHOE-SHAPED STEM WAS STILL INTACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
447699 M2A-MAGNUM PF CUP 54ODX48ID PROSTHESIS, HIP KWA BIOMET ORTHOPEDICS N/A 914730

Patients

Seq Age Sex Outcome Treatment
1 55 YR Hospitalization| R