LIFEVEST WCD 4000 SYSTEM
Report
- Report Number
- 3008642652-2014-01653
- Event Type
- Malfunction
- Date Received
- June 10, 2014
- Date of Event
- May 6, 2014
- Report Date
- May 27, 2014
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (WIRES EXPOSED, "ADD GEL" MESSAGES) HAS BEEN CONFIRMED. UPON INVESTIGATION, THE CABLE CONNECTING THE DISTRIBUTION NODE TO THE REAR THERAPY ELECTRODES WAS PULLED FROM THE STRAIN RELIEF, DAMAGING WIRES AND CAUSING THE REPORTED "ADD GEL" MESSAGES. THE ROOT CAUSE FOR THE STRAINED CABLE COULD NOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE STRAINED CABLE. THE PATIENT RECEIVED A REPLACEMENT ELECTRODE BELT.
A (B)(6) MALE PT CALLED ZOLL CUSTOMER SUPPORT TO REPORT THAT WIRES WERE EXPOSED ON HIS ELECTRODE BELT AND THE HE WAS RECEIVING "ADD GELL" MESSAGES WITHOUT PRIOR GEL RELEASE. THE PT WAS ISSUED A REPLACEMENT ELECTRODE BELT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 338358 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 41 YR |