FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3970473 · Received July 31, 2014

Report

Report Number
2531779-2014-21927
Event Type
Malfunction
Date Received
July 31, 2014
Report Date
July 26, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP #1 - DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 08/22/2014 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE PUMP¿S BLACK BOX DATA AND ALARM HISTORY REVEALED MULTIPLE LOW BATTERY WARNINGS. THE EVIDENCE OF FREQUENT LOW BATTERY ALARMS REVEALED EVIDENCE OF SHORT BATTERY LIFE. DURING TESTING, THE BATTERY COMPARTMENT WAS OBSERVED TO BE CRACKED. THE BATTERY CAP WAS ABLE TO FULLY SECURE ONTO THE PUMP. THE CURRENT DRAWS MEASURED WITHIN SPECIFICATIONS. THE PUMP WAS OPENED AND NO EVIDENCE OF MOISTURE OR LOOSE COMPONENTS WAS OBSERVED INSIDE THE PUMP. UNRELATED TO THE INITIAL COMPLAINT, THE DISPLAY SCREEN WAS DIM AND DISCOLORED.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (BATTERY LIFE) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
448064 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 52 YR