FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 3970447 · Received July 31, 2014

Report

Report Number
1416980-2014-24789
Event Type
Malfunction
Date Received
July 31, 2014
Date of Event
July 10, 2014
Report Date
July 10, 2014
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). AN ALARM INDICATIVE OF A POTENTIAL MALFUNCTION OF THE DISPOSABLE CASSETTE WAS IDENTIFIED. AS THE CASSETTE WAS NOT RETURNED AND THE LOT NUMBER IS UNKNOWN, A DEVICE ANALYSIS CANNOT BE COMPLETED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE HOME PATIENT (HP) HAD A SYSTEM ERROR (SE) 2240 (AIR IN LINE/SET) ON THE HOMECHOICE (HC) DEVICE DURING DWELL 3 OF 5 OF PERITONEAL DIALYSIS (PD) THERAPY, WHILE THE HP WAS CONNECTED. NOTHING WAS FOUND DURING TROUBLESHOOTING THAT WOULD HAVE CAUSED OR CONTRIBUTED TO THE ALARM. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) EXPLAINED THE ALARM AND ASSISTED THE HP IN CLEARING IT BY CYCLING THE POWER ON THE HC DEVICE. THE HP PLANNED TO COMPLETE THERAPY USING MANUAL SUPPLIES. THERE WAS NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE EVENT ASSOCIATED WITH THIS REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
447507 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 69 YR HOMECHOICE