HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Report
- Report Number
- 1416980-2014-24783
- Event Type
- Malfunction
- Date Received
- July 31, 2014
- Date of Event
- July 13, 2014
- Report Date
- July 13, 2014
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). AN ALARM INDICATIVE OF A POTENTIAL MALFUNCTION OF THE DISPOSABLE CASSETTE WAS IDENTIFIED. THE DEVICE WAS NOT RETURNED; THEREFORE, A DEVICE ANALYSIS COULD NOT BE COMPLETED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT A SYSTEM ERROR 2240 (AIR IN SET/LINE) ALARM OCCURRED ON THE HOMECHOICE (HC) DEVICE DURING DRAIN 1 OF 3. THE HOME PATIENT (HP) WAS CONNECTED AT THE TIME OF THE ALARM. DURING TROUBLESHOOTING, THERE WAS NOTHING UNUSUAL FOUND THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) EXPLAINED THE ALARM TO THE HP AND ASSISTED THE HP TO CLEAR THE ALARM. THE TSR REVIEWED PROPER PROCEDURES WITH THE HP. THE HP WOULD COMPLETE THERAPY USING ALL NEW SUPPLIES. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 448026 | HOMECHOICE AUTOMATED PD SET WITH CASSETTE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 52 YR | HOMECHOICE |