FDA Adverse Event Malfunction Summary report: N

MONOPOLAR CURVED SCISSORS INSTRUMENT

MDR report key: 3970331 · Received July 31, 2014

Report

Report Number
2955842-2014-04631
Event Type
Malfunction
Date Received
July 31, 2014
Date of Event
July 10, 2014
Report Date
July 22, 2014
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. FAILURE ANALYSIS WAS UNABLE TO CONFIRM THE CUSTOMER REPORTED COMPLAINT OF MONOPOLAR CORD BURNED THROUGH AT SITE OF ATTACHMENT. THE INSTRUMENT DID NOT SHOW ANY SIGNS OF ARCING. THE INSTRUMENT PASSED A CAUTERY TEST ON AN IN-HOUSE DA VINCI SYSTEM. FAILURE ANALYSIS INVESTIGATION ALSO FOUND THE DISTAL END OF THE INSTRUMENT'S MAIN TUBE HAD VARIOUS SCRATCH MARKS/ABRASIONS WITH LIGHT MATERIAL REMOVAL. THE SCRATCHES MEASURED APPROXIMATELY 0.053 - 0.171 IN LENGTH AND WERE NOT ALIGNED WITH THE TUBE AXIS. IN ADDITION, THE INSTRUMENT'S MAIN TUBE WAS CRACKED. THE TUBE HAD A SMALL FISSURE IN THE AXIAL DIRECTION. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT IN ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE VARIOUS SCRATCH MARKS ON THE MAIN TUBE WITH LIGHT MATERIAL REMOVAL FOUND DURING FAILURE ANALYSIS IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI HYSTERECTOMY SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE MONOPOLAR CORD BURNED THROUGH AT SITE OF ATTACHMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
448265 MONOPOLAR CURVED SCISSORS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420179-12 N10140422 975

Patients

Seq Age Sex Outcome Treatment
1