FDA Adverse Event Malfunction Summary report: N

MONOPOLAR CURVED SCISSORS INSTRUMENT

MDR report key: 3970308 · Received July 31, 2014

Report

Report Number
2955842-2014-04630
Event Type
Malfunction
Date Received
July 31, 2014
Date of Event
April 21, 2014
Report Date
July 22, 2014
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. FAILURE ANALYSIS WAS UNABLE TO CONFIRM THE CUSTOMER REPORTED COMPLAINT OF ARTICULATING END ROTATED DURING THE CASE. THERE WAS NO TROUBLE FOUND. THE INSTRUMENT WAS RETURNED WITH ZERO USES LEFT SO IT WAS NOT POSSIBLE TO TEST THE INSTRUMENT FOR ARTICULATION. VISUAL INSPECTION FOUND NO PROBLEMS WITH THE CABLES AT THE DISTAL END OR BACK END. THE INSTRUMENT WAS MANUALLY MOVED AND NO PROBLEMS WERE OBSERVED. FAILURE ANALYSIS INVESTIGATION FOUND THE INSTRUMENT'S TUBE EXTENSION EXHIBITED REMOVED PAD PRINTING AND VARIOUS SCRATCH MARKS. THE REMOVED PAD PRINTING AREA MEASURED APPROXIMATELY 0.299 X 0.058 - 0.120 X 0.068. FAILURE ANALYSIS CONCLUDED THAT THE PAD PRINTING DAMAGE MAY BE DUE TO MISHANDLING/MISUSE. NO OTHER DAMAGE WAS FOUND. THE INSTRUMENTS AND ACCESSORIES USER MANUAL SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. THE CUSTOMER REPORTED COMPLAINT DOES NOT IN ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE PAD PRINTING DAMAGE IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI SALPINGECTOMY SURGICAL PROCEDURE, THE ARTICULATING END ON THE MONOPOLAR CURVED SCISSORS INSTRUMENT ROTATED AND NEEDED TO BE RESET THREE TIMES. THERE WAS NO REPORT OF FRAGMENTS FALLING INTO A PATIENT. THERE WAS NO PATIENT HARM, ADVERSE OUTCOME OR INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
448224 MONOPOLAR CURVED SCISSORS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420179-12 M13130808 508

Patients

Seq Age Sex Outcome Treatment
1 25 YR