HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Report
- Report Number
- 1416980-2014-24781
- Event Type
- Malfunction
- Date Received
- July 31, 2014
- Date of Event
- July 13, 2014
- Report Date
- July 13, 2014
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). AS THE CASSETTE WAS NOT RETURNED AND THE LOT NUMBER IS UNKNOWN, A DEVICE ANALYSIS CANNOT BE COMPLETED. USE ERRORS AND PROPER USER INSTRUCTIONS ARE ADDRESSED IN ¿THE HOMECHOICE AND HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE¿, WHICH IS SHIPPED WITH EVERY HOMECHOICE DEVICE. THE GUIDE PROVIDES STEP-BY-STEP INSTRUCTIONS FOR PROPERLY PRIMING THE DISPOSABLE SET AND WARNS THE USER NOT TO CONNECT TO THE PATIENT LINE UNLESS THE FLUID LEVEL IS AT OR NEAR THE CONNECTOR AT THE END OF THE DISPOSABLE SET PATIENT LINE. THE GUIDE ALSO INSTRUCTS THE USER TO VERIFY THAT THE PATIENT LINE IS PROPERLY PRIMED AND PROVIDES STEP-BY-STEP INSTRUCTIONS FOR PROPERLY REPRIMING THE PATIENT LINE. A FORMAL REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
DURING A REQUEST FOR THERAPY ASSISTANCE ON THE HOMECHOICE (HC) MACHINE, IT WAS REPORTED THAT THE HOME PATIENT (HP) CONNECTED TO THE SETUP BEFORE ¿CONNECT YOURSELF¿ WAS DISPLAYED. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) EXPLAINED TO THE HP THAT THEY SHOULD NOT BE CONNECTED WHILE THE HC DEVICE IS PRIMING. THE TSR ADVISED THE HP TO START THERAPY OVER WITH NEW SUPPLIES. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 447100 | HOMECHOICE AUTOMATED PD SET WITH CASSETTE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR | HOMECHOICE |