LIGASURE ADVANCE PISTOL GRIP
Report
- Report Number
- 1717344-2014-00379
- Event Type
- Malfunction
- Date Received
- April 29, 2014
- Report Date
- April 25, 2014
- Manufacturer
- COVIDIEN LP
- Product Code
- GEI
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- HI, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). PART OF THE CLEAR INSULATION WAS MISSING. THE DEVICE FAILED HIPOT TESTING. THE IFU STATES TO PREVENT DAMAGE TO THE FLEXIBLE INSULATION PROXIMAL TO THE JAWS, CONFIRM THE HANDLE IS FULLY CLOSED PRIOR TO INSERTION INTO AND EXTRACTION FROM THE CANNULA. USE THE APPROPRIATELY SIZED CANNULA TO ALLOW FOR EASY INSERTION AND EXTRACTION OF THE INSTRUMENT. FAILURE TO DO SO MAY IMPACT THE INTEGRITY OF THE FLEXIBLE INSULATION. CANNULAS WITH HARD, NON-BEVELED OPENINGS MAY CAUSE THE FLEXIBLE INSULATION TO RETRACT, WHICH MAY COMPROMISE THE INSULATION. IF RETRACTION OCCURS, THE INSTRUMENT MUST BE DISCARDED. DO NOT ATTEMPT TO CLEAN THE FLEXIBLE INSULATION. CLEANING MAY DAMAGE INSULATION. THE JAWS WERE NOT JAMMED SHUT. THE JAWS OPENED AND CLOSED NORMALLY WHEN THE HANDLE WAS ACTIVATED.
THE CUSTOMER REPORTED THAT THE DEVICE WAS NOT OPENING PROPERLY DURING A LAPAROSCOPIC LOW ANTERIOR RESECTION. THERE WAS NO INJURY TO THE PATIENT. THE DEVICE WAS RETURNED TO COVIDIEN AND VISUAL INSPECTION DISCOVERED THAT PART OF THE CLEAR INSULATION WAS NO LONGER ATTACHED. THE SITE CONFIRMED THAT THE INSULATION DID NOT FALL INTO THE PATIENT ACTIVITY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 256114 | LIGASURE ADVANCE PISTOL GRIP | LIGASURE VESSEL SEALING SYSTEM | GEI | COVIDIEN LP | 32900145X |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |